One-Stop Reference for all Quality Assurance Process Activity

Ensuring Tablet Integrity: A Comprehensive Guide to Calibration and Specification in Tablet Friability Testing

Contributing Author:   Prasad Harichandra Submitted Date: 07 Feb 2008

This article delves into the crucial aspects of Tablet Friability Test Calibration and Specification, shedding light on the processes that ensure the accuracy and reliability of this essential quality control measure in pharmaceutical manufacturing. Covering calibration procedures and specification standards, the guide provides pharmaceutical professionals with insights into maintaining the integrity of tablet formulations. Understanding and implementing precise calibration and specifications is vital for upholding product quality and meeting stringent regulatory requirements.

The tablet friability test is an important quality control test in the pharmaceutical industry used to assess the mechanical strength and durability of tablets. It measures the tendency of tablets to chip, crack, or break when subjected to mechanical stress during handling, packaging, and transport. Proper calibration and specification of the friability test apparatus are essential to ensure accurate and reliable results. Here's an overview:

Calibration of the Friability Test Apparatus:

  1. Frequency: The friability test apparatus should be calibrated regularly, as specified in the laboratory's quality control procedures and in accordance with regulatory requirements. The frequency of calibration may vary but is often done annually or semi-annually.

  2. Calibration Standards: Use calibrated reference tablets with known friability values or a standard calibration kit provided by the equipment manufacturer for calibration. These reference tablets should closely match the characteristics of the tablets being tested.

  3. Calibration Procedure:

    • Place the reference tablets in the friability drum or basket of the apparatus.
    • Set the appropriate test parameters, including rotation speed and test duration, as per the official test method or product-specific requirements.
    • Run the calibration cycle.
    • Record the results and compare them to the known friability values of the reference tablets or the expected values provided by the calibration kit.
  4. Adjustment: If the results of the calibration are outside acceptable limits, make necessary adjustments to the friability test apparatus to bring it into calibration. This may involve recalibrating or repairing the equipment.

Specification for Tablet Friability:

  1. Acceptance Criteria: The specification for tablet friability should define the acceptable level of tablet weight loss during the test. This is usually expressed as a percentage.

  2. Regulatory Standards: The acceptable friability levels are often specified in pharmacopeial standards, such as the United States Pharmacopeia (USP) or the European Pharmacopoeia (Ph. Eur.), and may vary depending on the type of tablet and its intended use.

  3. Typical Specification: A typical specification for tablet friability may be set at 0.5% or lower. This means that the total weight loss of tablets during the friability test should not exceed 0.5% of their initial weight.

  4. Product-Specific Requirements: In some cases, specific pharmaceutical products may have tighter friability specifications based on their formulation, dosage form, or intended use. Manufacturers should establish and adhere to these specifications as needed.

  5. Repeat Testing: If tablets fail the friability test, repeat the test on additional samples to confirm the results. If multiple tests consistently fail, it may indicate a problem with the tablet formulation or manufacturing process.

  6. Documentation: Maintain comprehensive records of the friability test results, including the date, batch or lot number, test parameters, and any deviations from the specifications.

  7. Investigation and Corrective Action: If tablets consistently fail the friability test, investigate the root cause of the issue, and take appropriate corrective actions. This may involve process adjustments or changes to the tablet formulation.

The tablet friability test and its associated calibration and specification are critical for ensuring that pharmaceutical tablets can withstand the rigors of handling and transportation without excessive damage. Adherence to specified limits, accurate calibration of equipment, and thorough documentation are essential for maintaining product quality and patient safety.


Keywords:#TabletFriabilityTest #QualityControl #PharmaceuticalManufacturing #CalibrationStandards #SpecificationGuidelines #PharmaCompliance #TabletIntegrity #RegulatoryStandards #QualityAssurance #DrugQuality #TabletTesting


Demonstation Video

Please check out the demonstation.


Request Information

reach out to us for learn more about our service offerings.


Please complete our request information form and our consulting representative shall get in touch with you shortly.